Designed specifically for Aquaculture Systems, the YSI 5400 and AquaManager® Software integrate process control, alarming, and data management into one product. Simple enough to monitor one tank, powerful enough to manage a full scale farming operation from anywhere in the world. Simultaneously measure dissolved oxygen in multiple tanks or ponds.
- Overview
- Specifications
Buffer Recognition Yes Certifications CE, NEMA 4X, Made in USA, CUL, ETL, RoHS, ULC, IP-65 Connectivity / Communications RS-485 and RS-232 standard, Ethernet and wireless optional Data Management AquaManager™ Software integrates process control, alarming, and data management into one product Desktop Software Compatible AquaManager®, Windows 2000/XP/Vista/Windows 7 operating systems Display Back-lit graphic LCD (320 x 240) with timer, extended temperature range LCD Equipment used with AquaManager Desktop Software, 5420-4, or 30, 5421-4, 20, 20 or 30, 10 GLP Compliance Yes Logging Capabilities Yes Measurement Method / Channels pH, 0 to 5.0 VDC or 4-20 mA) 8 each 10 amp 115 VAC (5 amp 230 VAC) relay, 4 DO probe inputs 4 user-configurable inputs for temp (0 to 1.0 VDC, digital (0 to 1.0 VDC, photo period, etc. 2 user-configurable inputs, TGP, ORP, 0 to 5.0 VDC or 4-20 mA) or DO Memory Non-volatile, 1000 data, 1250 relay/alarm, 1000 entries/change and 500 GLP records Monitoring Yes Operating Temperature -15 to +70°C (5 to 158°F), direct outside conditions with NEMA 4X enclosure Parameters Measured Dissolved Oxygen % Saturation, Dissolved Oxygen mg/L, Temperature (optional) Power 12 V DC, mobile live haul or anytime AC power not available User Calibratable Yes User Replaceable Yes Warranty 2-years instrument; 1-year probes and cable Waterproof Yes Specification Timers 4 timers - 10 events cascading events start/duration
1 feed timer channel (feeder not included); Feed Smart software manages feed delivery, user-selectable inputs for number of daily feedings, daily feed weights, total biomass, FCRS and parameter controlOutput Relays 8 each 10 amp 115 VAC (5 amp 230 VAC) relays for parameter, control and/or alarm
E-mail and/or SMS alarming up to 3 e-mails with Ethernet module or up to 10 e-mails with AquaManagerField Upgradeable Software Yes; flash based Height 6.5 in, 16.5 cm Length 8.5 in, 21.6 cm Depth 4.75 in, 12.1 cm Parameter
(units)Range Accuracy Resolution Dissolved Oxygen*
(% saturation)0 to 500% 0 to 200%: ±2% or ±0.2 mg/L, whichever is greater;
200 to 500%: ±6% of reading0.1% air saturation Dissolved Oxygen*
(mg/L, ppm)0 to 60 mg/L 0 to 20 mg/L: ± 0.2 mg/L or ±2% of reading, whichever is greater;
20 to 60 mg/L: ±6% of reading0 to 10 mg/L: 0.01 mg/L
10 – 60 mg/L: 0.1 mg/LTemperature
(°C, °F)0 to +45°C (32 to 113°F) ±0.2°C 0.1°C *Galvanic dissolved oxygen sensor. Response Time: 95% of reading in 17 seconds for 2.0 mil PE membrane; 95% of reading in 36 seconds for 2.0 mil Teflon® membrane.
Specifications indicate typical performance and are subject to change.