Designed specifically for Aquaculture Systems, the YSI 5200A and AquaManager™ Software integrate process control, alarming, and data management into one product. Simple enough to monitor one tank, powerful enough to manage a full scale farming operation from anywhere in the world. Simultaneously measure dissolved oxygen, pH, conductivity, ORP, salinity and temperature.
- Overview
- Specifications
Buffer Recognition Yes Certifications RoHS, IP-65, ETL, CUL, NEMA 4X, CE, Made in USA, ULC Connectivity / Communications Ethernet and wireless optional. Output Relays: 4 each 10 amp 115 VAC (5 amp 230 VAC) relays for parameter, control and/or alarm E-mail and/or SMS alarming up to 3 e-mails with Ethernet module or up to 10 e-mails with AquaMana, RS-485 and RS-232 standard Data Management AquaManager™ Software integrates process control, alarming, and data management into one product Desktop Software Compatible Windows 2000/XP/Vista/Windows 7 operating systems, AquaManager® Display Back-lit graphic LCD (320 x 240) with timer, extended temperature range LCD Equipment used with or 30, 10, 5561-4, 20, or 20, AquaManager Desktop Software, 5562-4 GLP Compliance Yes Logging Capabilities Yes Measurement Method / Channels DO, pH, ORP, conductivity, temperature, 2 user-configurable inputs, digital (0 to 1.0 VDC, 0 to 5.0 VDC or 4-20 mA) Timers: 4 timers - 10 events cascading events start/duration 1 feed timer channel (feeder not included), Feed Smart software manages feed Memory Non-volatile, 1000 data, 1250 relay/alarm, 1000 entries/change and 500 GLP records Monitoring Yes Multiparameter Yes Operating Temperature -15 to +70°C (5 to 158°F), direct outside conditions with NEMA 4X enclosure Parameters Measured Dissolved Oxygen (electrochemical), pH,ORP/Redox, Conductivity, Specific Conductance, Salinity, Total Dissolved Solids (TDS), Temperature Power 12 V DC, mobile live haul or anytime AC power not available User Calibratable Yes User Replaceable Yes Warranty 2-years instrument; 1-year probes and cable Waterproof Yes Specification Timers 4 timers - 10 events cascading events start/duration
1 feed timer channel (feeder not included); Feed Smart software manages feed delivery, user-selectable inputs for number of daily feedings, daily feed weights, total biomass, FCRS and parameter controlField Upgradeable Software Yes; flash based Height 6.5 in, 16.5 cm Length 8.5 in, 21.6 cm Depth 4.75 in, 12.1 cm Parameter
(units)Range Accuracy Resolution Dissolved Oxygen*
(% saturation)0 to 500% 0 to 200%: ±2% or ±0.2 mg/L, whichever is greater;
200 to 500%: ±6% of reading0.1% air saturation Dissolved Oxygen*
(mg/L, ppm)0 to 60 mg/L 0 to 20 mg/L: ± 0.2 mg/L or ±2% of reading, whichever is greater;
20 to 60 mg/L: ±6% of reading0 to 10 mg/L: 0.01 mg/L
10 – 60 mg/L: 0.1 mg/LTemperature
(°C, °F)0 to +45°C (32 to 113°F) ±0.2°C 0.1°C Conductivity
(µS, mS)0 to 200 µS;
0 to 2000 µS
0 to 20 mS;
0 to 200 mS0 to 100 mS: ±0.5% of range
100 to 200 mS: ±1.0% of reading0.1 µS
1 µS
0.01 mS
0.1 mS
(range dependent)Salinity
(ppt)0 to 80 ppt (0 to 50 ppt when entered manually) ±2% or ±0.1 ppt,
whichever is greater0.1 ppt pH
(pH units)0 to 14 units ±0.2 units 0.01 units ORP
(mV)-2000 to +2000 mV ±20 mV 1 mV *Polarographic dissolved oxygen sensor
Specifications indicate typical performance and are subject to change.