※The new FN-325 can not be used with SWT-7000, SWT-7000Ⅱ series but with SWT-7000Ⅲsereis only | ||||||
Specifications | ||||||
Probe type : Dual FN-325 (exclusive for SWT-7000Ⅲ series) Measuring method : Electromagnetic/Eddy current dual use Measuring range : Ferrous:0~3.00mm, Nonferrous:0~2.50mm Substrate detection : Automatic or manual switching Display resolutions : 1μm:0~999μm By switching, 0.1μm:0~400μm, 0.5μm:400~500μm, 0.01mm:(ferrous 1.00~3.00mm) , (nonferrous 1.00~2.50mm) Accuracy (place the probe perpendicularly : Ferrous &nonferrous dual use to the flat face) 0~100μm:±1μm or ±2% the read value (ferrous)101μm~3.00mm:±2% (nonferrous)101μm~2.50mm:±2% Probe : One point contact constant pressure type, with V cut φ13×52mm, 72g Option : V type probe adaptors (3 types: less Φ5, Φ5~10, Φ10~20 ) Accessories : Calibration standards (plastic foils) Zero plate (ferrous, nonferrous dual) Measuring objectives : Fe substrate: coating, lining, thermal spraying, plating on magnetic metal like iron, steel (except electrolyte nickel plating) NFe substrate: insulated films on non-magnetic metal like aluminum, copper |