handheld-multiparameter-instrument-with-digital-sensors-2.png

Handheld Multiparameter Instrument with Digital Sensors

The YSI ProDSS (digital sampling system) handheld multiparameter meter provides extreme flexibility for the measurement of a variety of combinations for dissolved oxygen, turbidity, conductivity, specific conductance, salinity, resistivity, total dissolved solids (TDS), pH, ORP, pH/ORP combination, ammonium (ammonia), nitrate, chloride and temperature along with depth.

Some sensors are EPA approved for drinking water and wastewater compliance reporting.

Instrument Only.  Cables, probes/sensors, and accessories sold separately.

 

  • Overview 
  • Specifications 
    Benchtop Yes
    Buffer Recognition Yes
    Certifications WEEE, CE, FCC, IP-67, Assembled in USA, 1-meter drop test, RoHS, UN Part III Section 38.3 test methods for lithium-ion batteries (class 9)
    Connectivity / Communications Built-in micro USB port
    Connector MS (military spec) waterproof with quarter-turn bayonet lock (unsurpassed in regards to durability)
    Data Management KorDSS desktop PC software (included), >100,000 data sets
    Desktop Software Compatible KorDSS desktop PC software (included)
    Digital / Digital Sensors Yes
    Flow Cell Yes, flow cell is available
    GLP Compliance detailed GLP information is stored and is available to view, download or print, Yes
    GPS Optional internal GPS, coordinates are stored with measurement data and site lists
    Graphic Display backlit keypad, Color, LCD graphic display
    Languages French, Spanish, Japanese, Portuguese, German, Italian, Norwegian, English, Chinese (Simplified & Traditional)
    Logging Capabilities Single point or Continuous, Yes
    Measurement Method / Channels Digital
    Memory >100,000 data sets, 400 GLP files
    Multiparameter Yes
    Operating Temperature 0 to 50°C (32 to 122°F)
    Parameters Measured Dissolved Oxygen (%), Dissolved Oxygen (mg/L), Temperature, Turbidity, Conductivity, Specific Conductance, Resistivity, Salinity, Water Density, pH, ORP, Ammonium, Nitrate, Chloride, Total Dissolved Solids (TDS), Total Suspended Solids (TSS), Depth, Barometer
    Power Rechargeable lithium-ion battery pack providing ~48 hours of battery life (instrument only with backlight off)
    Sampling Yes
    Security / Password Protection / UserID Yes
    Size

    Instrument: 8.3 cm width x 21.6 cm length x 5.6 cm depth (3.27 in x 8.5 in x 2.21 in)
    Bulkhead with sensors, without depth (no guard): 33.3 cm (13.11 in) length - the length with the ISE sensors is 34.04 cm (13.4 in)
    Bulkhead with sensors, with depth (no guard): 35.84 cm (14.11 in) length - the length with the ISE sensors is 36.58 cm (14.4 in)
    Bulkhead with guard, without depth: 42.82 cm (16.86 in) length and 4.75 cm (1.87 in) outer diameter
    Bulkhead with guard, with depth: 45.36 cm (17.86 in) length and 4.75 cm (1.87 in) outer diameter

    Smart Sensors / Ports Yes, 4 universal sensor ports on each cable, can accept any ProDSS sensor
    Storage Temperature 0 to 45°C (32 to 113°F) with battery installed0 to 60°C (32 to 140°F) without battery installed
    User Calibratable Yes
    User Replaceable Yes
    Warranty 3-year instrument; 2-year bulkhead, cable assembly, and sensors; 1-year pH and pH/ORP sensor modules, ODO (opticalDO) sensor caps, and Li-Ion battery pack; 6-months ammonium, nitrate, and chloride sensor modules
    Waterproof IP-67 (even with the battery cover off), floats

    ProDSS Instrument General Specifications

    Specification  
    Barometer Built-in barometer
    Certifications RoHS, CE, WEEE, FCC, IP-67, 1-meter drop test, UN Part III Section 38.3 test methods for lithium-ion batteries (class 9), Assembled in USA
    Connectivity Built-in micro USB port
    Connector MS (military spec) waterproof with quarter-turn bayonet lock (unsurpassed in regards to durability)
    Data Management

    KorDSS desktop PC software (included)

    Data Memory >100,000 data sets; 400 GLP files
    Dimensions 8.3 cm width x 21.6 cm length x 5.6 cm depth (3.25 in x 8.5 in x 2.21 in)
    Display Color, LCD graphic display; backlit keypad
    Cables With or without depth sensors in lengths of 1, 4, 10, 20, 30, 40, 50, 60, 70, 80, 90 and100-meters; 10, 20 and 30-meter cables include a cable management kit; all cables 10-meters and longer include a 1 lb weight
    Flow Cell Yes; flow cell is available
    GLP Compliance Yes; detailed GLP information is stored and is available to view, download or print
    Languages English, Spanish, German, French, Italian, Norwegian, Portuguese, Japanese, Chinese (Simplified & Traditional)
    Logging Modes Single point or Continuous
    Operating Temperature 0 to 50°C (32 to 122°F)
    Power Rechargeable lithium-ion battery pack providing ~48 hours of battery life (instrument only with backlight off)
    Storage Temperature 0 to 45°C (32 to 113°F) with battery installed
    0 to 60°C (32 to 140°F) without battery installed
    Sites and Data ID 100 user-defined sites and 100 user-defined data ID tags
    Warranty 3-year instrument; 2-year bulkhead, cable assembly, and sensors; 1-year pH and pH/ORP sensor modules, ODO (opticalDO) sensor caps, and Li-Ion battery pack; 6-months ammonium, nitrate, and chloride sensor modules
    Waterproof IP-67 (even with the battery cover off); floats
    Meter Weight with Batteries 567 grams (1.25 lbs)
    Bulkead Weight with 4 Sensors 680 grams (1.5 lbs) with sensor guard on; 589 grams (1.3 lbs) without sensor guard
    GPS Optional internal GPS; coordinates are stored with measurement data and site lists
    Sensor Ports 4 universal sensor ports on each cable; can accept any ProDSS sensor

     

    ProDSS System Specifications (Instrument, Cables & Sensors)

    Parameter
    (units)
    Sensor Type Range Accuracy Resolution Calibration Maximum Depth
    Dissolved Oxygen
    (% saturation)
    Optical Luminescence - Lifetime Method 0 to 500% 0 to 200% (±1% of reading or 1% air saturation, whichever is greater)
    200% to 500% (±8% of reading)
    0.1% or 1% air saturation (user selectable) 1 or 2 points 100 m
    Dissolved Oxygen(mg/L, ppm)
    temp comp range -5 to 50°C
    Optical Luminescence - Lifetime Method 0 to 50 mg/L 0 to 20 mg/L
    (±0.1 mg/L or 1% of reading, whichever is greater)
    20 – 50 mg/L (±8% of reading)
    0.1 or 0.01 mg/L (user selectable) 1 or 2 points (user selectable) 100 m
    Temperature
    (°C, °F, K)
    Thermistor; Combination Sensor with Conductivity -5 to 70°C (23 to 158°F) ±0.2°C 0.1°C or 0.1°F (user selectable) None 100 m
    Turbidity
    (FNU, NTU)
    Nephelometric - Optical, 90° Scatter 0 to 4000 FNU 0 to 999 (0.3 or ±2% of reading, whichever is greater)
    1000 to 4000 (±5% of reading)
    0.1 FNU 1, 2, or 3 points (user selectable) -
    Conductivity*
    (µS, mS)
    Four Nickel Electrode Cell 0 to 200 mS/cm 0 to 100 mS/cm (±0.5% of reading or 0.001 mS/cm, whichever is greater)
    100 to 200 mS/cm (±1% of reading)
    0.001, 0.01 or 0.1 µS/cm (range dependent) 1 point 100 m
    Specific Conductance
    (µS, mS)
    Calculated from Conductivity and Temperature 0 to 200 mS/cm ±0.5% of reading or 0.001 mS/cm, whichever is greater 0.001, 0.01, 0.1 mS/cm 1 point -
    Resistivity
    (ohm-cm, kohm-cm, Mohm-cm)
    Calculated from Conductivity and Temperature 0 to 2 Mohms ±0.1% Full Scale 0.001, 0.01, 0.1 ohms - -
    Salinity
    (ppt, PSU)
    Calculated from Conductivity and Temperature 0 to 70 ppt ±1.0% of reading or ±0.1 ppt, whichever is greater 0.01 ppt 1 point -
    Water Density
    (Sigma, Sigma T)
    Sigma is Calculated from Salinity, Temperature, and Pressure
    Sigma T is Calculated from Salinity and Temperature
    0.0 to 50.0 sigma, sigma T - 0.1 sigma or sigma T - -
    pH
    (mV, pH units)
    Glass Bulb Combination Electrode; Ag/AgCl Reference Gel 0 to 14 units ±0.2 units 0.01 units 1, 2, or 3 points (user selectable) 100 m
    ORP
    (mV)
    Platinum Button; Ag/AgCl Reference --1999 to +1999 mV ±20 mV 0.1 mV 1 point 100 m
    Ammonium**
    (NH4-N mg/L, NH4-N mV)
    ammonia with pH sensor
    Ion Selective Electrode 0 to 200 mg/L NH4-N ±10% of reading or 2 mg/L, whichever is greater 0.01 mg/L 1, 2, or 3 point
    (user selectable)
    20 m
    Ammonia***
    (NH3-N mg/L)
    Calculated from Ammonium, Temperature, Salinity and pH 0 to 200 mg/L NH4-N - 0.01 mg/L - -
    Nitrate**
    (NO3-N mg/L, NO3-N mV)
    Ion Selective Electrode 0 to 200 mg/L NO3-N ±10% of reading or 2 mg/L, whichever is greater 0.01 mg/L 1, 2, or 3 point
    (user selectable)
    20 m
    Chloride** Ion Selective Electrode 0 to 18000 mg/L Cl ±15% of reading or 5 mg/L, whichever is greater 0.01 mg/L 1, 2, or 3 point
    (user selectable)
    20 m
    Total Dissolved Solids (TDS)
    (mg/L, kg/L, g/L)
    Calculated from Conductivity and Temperature 0 to 100 g/L User-Selectable TDS Multiplier (0.30 to 1.00; 0.65 default) 0.001, 0.01, 0.1g/L - -
    Total Suspended Solids (TSS)
    (mg/L)
    Correlated from Turbidity Field Measurements and Lab TSS Measurements from Grab Samples 0 to 30000 mg/L Calculated from Turbidity and User-Entered Correlation Points 0.01, 0.1 mg/L - -
    Depth
    (m, ft)
    Pressure Transducer 0 to 100 m (0 to 328 ft) ±0.004 m (±0.013 ft) for 1, 4 and 10-m cables
    ±0.04 m (±0.13 ft) for cables 20-m and longer
    0.001 m or 0.01 ft 1 point -
    Barometer
    (mmHg, inHg, mbar, psi, kPa, ATM)
    Piezoresistive 375 to 825 mmHg ±1.5 mmHg from 0 to 50°C 0.1 mmHg 1 point -

    * Derived parameters can include resistivity, salinity, specific conductance, and total dissolved solids
    **ISE sensors for freshwater only; 20-meter maximum depth