turbiscat.png


In-Line Turbidimeter
The TurBiScat monitors turbidity in liquids, optional also the (beer)colour in EBC. The combination of Hastelloy and sapphire in a sealless design allows operating the unit in practically all process applications. Turbidity measurement is done with colour compensation. Window fouling is monitored, compensated and can be checked using an anti-fouling control. Calibration is easily done using a secondary reference standard

SPECIFICATION

Wavelength turbidity:   LED 650 nm
Wavelength color (optional):   LED 430 nm
Measuring principle:   90°/25° scattered light measurement
Resolution:   < 0.001 EBC of full scale
Reproducibility:   ±1 % @ 0 .. 2 EBC
                           ±2 % @ 2 .. 100 EBC (90°)
                           ±3 % @ 2 .. 100 EBC (25°)
                           ±10 % @ 100 .. 1,000 EBC
Repeatability:   0.001 EBC or ±0.25% of full scale
Reaction time:   less than 2 s (step response - limit value switch)
Enclosure:   Stainless steel 1.4301
Protection class:   IP 66
Supply voltage:   9 .. 30 VDC, 3 W (of control unit)
Dimensions:   Ø100.5 x 118.5 mm
Weight:   Approx. 2.3 kg
Sample temperature:   -10 .. +100 °C
                                    120 °C max. 2 h
                                    150 °C max. 1 h
                                    180° C max. with cooling (optional)
Sample pressure:   2 MPa (20 bar) @ 180 °C
                              4 MPa (40 bar) @ 100 °C
Connections:   DN 40 .. DN 125
                        1½" .. 6" Varivent®
Ambient environment:   -10 .. +50 °C
                                      0 .. 100 % relative humidity
Nominal range color:   0 .. 50 EBC
Nominal range turbidity:   0 .. 1,000 EBC
                                          0 .. 4,000 NTU
Linearity:   ±0.5% full scale @ 0 .. 2 EBC
Warm-up time:   < 3 min.
Temperature stability:   < -0.15 %K-1 of full scale
Cleaning:   CIP/SIP compatible up to 120 °C
Material sensor head:   Hastelloy C-22
Windows material:   Sapphire
Control Units:   SICON, SICON M